Résumé

Since the beginning of the 1990s, the CEA is being involved in the development of compact gamma cameras based on a pinhole collimator, a scintillator and an intensified CCD. In collaboration with the Kurchatov Institute, the CEA has recently developed a miniature coded mask, which significantly improves sensitivity and angular resolution. Laboratory tests have demonstrated excellent performance. They are summarized, accompanied with on-site test results. We also present preliminary results of Medipix2/CdTe, a semiconductor pixel detector.
Fichier principal
Vignette du fichier
2006_NIM-A-563_233-237_2006_Postprint-auth.pdf (835.7 Ko) Télécharger le fichier
Origine Fichiers produits par l'(les) auteur(s)
licence

Dates et versions

hal-04721864 , version 1 (29-10-2024)

Licence

Identifiants

Citer

Olivier Gal, Mehdi Gmar, Oleg Ivanov, Frédéric Lainé, Fabrice Lamadie, et al.. Development of a portable gamma camera with coded aperture. Nuclear Instruments and Methods in Physics Research Section A: Accelerators, Spectrometers, Detectors and Associated Equipment, 2006, 563 (1), pp.233-237. ⟨10.1016/j.nima.2006.01.119⟩. ⟨hal-04721864⟩