Spatial resolution in depth for time-resolved diffuse optical tomography using short source-detector separations

Résumé

Diffuse optical tomography for medical applications can require probes with small dimensions involving short source-detector separations. Even though this configuration is seen at first as a constraint due to the challenge of depth sensitivity, we show here that it can potentially be an asset for spatial resolution in depth. By comparing two fiber optic probes on a test object, we first show with simulations that short source-detector separations improve the spatial resolution down to a limit depth. We then confirm these results in an experimental study with a state-of-the-art setup involving a fast-gated single-photon avalanche diode allowing maximum depth sensitivity. We conclude that short source-detector separations are an option to consider for the design of probes so as to improve image quality for diffuse optical tomography in reflectance.
Fichier principal
Vignette du fichier
main.pdf (1.28 Mo) Télécharger le fichier
Origine Publication financée par une institution
licence

Dates et versions

hal-01131904 , version 1 (06-01-2025)

Licence

Identifiants

Citer

Agathe Puszka, Laura Di Sieno, Alberto Dalla Mora, Antonio Pifferi, Davide Contini, et al.. Spatial resolution in depth for time-resolved diffuse optical tomography using short source-detector separations. Biomedical optics express, 2015, 6 (1), pp.1-10. ⟨10.1364/BOE.6.000001⟩. ⟨hal-01131904⟩
37 Consultations
0 Téléchargements

Altmetric

Partager

More