Comparison of digital signal processing modules in gamma-ray spectrometry - CEA - Commissariat à l’énergie atomique et aux énergies alternatives Access content directly
Journal Articles Applied Radiation and Isotopes Year : 2014

Comparison of digital signal processing modules in gamma-ray spectrometry

Abstract

Commercial digital signal-processing modules have been tested for their applicability to gamma-ray spectrometry. The tests were based on the same n-type high purity germanium detector. The spectrum quality was studied in terms of energy resolution and peak area versus shaping parameters, using a Eu-152 point source. The stability of a reference peak count rate versus the total count rate was also examined. The reliability of the quantitative results is discussed for their use in measurement at the metrological level.
No file

Dates and versions

cea-01834192 , version 1 (10-07-2018)

Identifiers

Cite

Marie-Christine Lépy, Ousmane Ibrahima Cissé, Sylvie Pierre. Comparison of digital signal processing modules in gamma-ray spectrometry. Applied Radiation and Isotopes, 2014, 87, pp.402 - 406. ⟨10.1016/j.apradiso.2013.11.015⟩. ⟨cea-01834192⟩
22 View
0 Download

Altmetric

Share

Gmail Facebook Twitter LinkedIn More